- Overview
- Specifications
- Description
- Applications
- Feature
- Troubleshooting
- FAQ
- Recommended Products
Overview
Place of Origin: |
Sweden |
Brand Name: |
ABB |
Model Number: |
SCC-F 23212-0-110310 |
Packaging Details: |
Original new Factory Sealed |
Delivery Time: |
5-7 days |
Payment Terms: |
T/T |
Supply Ability: |
In stock |
Specifications
|
Parameter |
Value |
|
Model |
SCC-F 23212-0-110310 |
|
Product Type |
Sample Gas Feed Unit |
|
Dimensions |
26.4 x 36 x 27.4 cm |
|
Weight |
11.44 kg |
|
Mounting Type |
19" plug-in unit / wall mounting (via brackets) |
|
Gas Paths |
2 independent gas paths (with/without condensate pump) |
|
Gas Line |
Corrosion-proof, metal-free |
|
Integrated Functions |
Flow and condensate switching amplifier (built-in) |
|
Protection Rating |
IP20 (EN 60529) |
|
Operating Temperature |
+5°C to +45°C |
|
Storage/Transport Temperature |
-25°C to +60°C |
|
Relative Humidity |
Max 95% (non-condensing) |
|
Power Supply |
115 V AC / 230 V AC, 50/60 Hz |
|
Power Consumption |
Max 185 VA |
|
Sample Gas Pressure |
70–105 kPa (0.7–1.05 bar abs) |
|
Sample Gas Flow |
Gas paths 1 & 2: max 100 l/h; bypass max 200 l/h |
|
Sample Gas Inlet Temperature |
+10°C to +50°C |
|
Dew Point Requirement |
At least 5°C below lowest ambient temperature in gas path |
Description
ABB SCC-F 23212-0-110310 is a sample gas feed unit, which transports the sample gas to the analyzer and controls flow and condensate in two independent gas paths. A digital communication port connects the unit with the controller board of Advance Optima analyzers to integrate status and alarm signals from the sample handling.
Applications
Thermal power plants: Emission monitoring in flue gas desulfurization and denitrification systems.
Cement processing plants: Real-time analysis of kiln tail and process gas composition.
Petrochemical plants: Sampling and transportation of flammable or corrosive gases in production processes.
Iron and steel smelting plants: Fine-grained composition measurement of coke oven gas or converter gas.
Waste incineration plants: Strict pre-sampling treatment to ensure flue gas emissions meet standards.
Feature
Complete sample gas conditioning in connection with the SCC-C sample gas cooler
Flow rate display and adjustment Rotameter 10 to 100 l/h with needle valve (metal-free)
Service-friendly due to easily accessible modules
Six pneumatic versions
Two seperate gas paths with additional sample pump
Full integration into AO2000 system with optional interface card
Troubleshooting
1.Feed unit not working
Power supply interrupted: Reconnect the power supply.
Fuse blown: Replace fuse (5x20 mm T6.3H250V).
Pump motor blocked: Remove blockage.
Defective pump: Replace pump.
Defective diaphragm: Replace diaphragm.
2.Drops of condensate in the condensate monitor or flow monitor (liquid alarm)
Cause:
Condensate being produced by the gas analysis system
Fluid from the process penetrating
Sample gas cooler’s condensate collecting vessel full
Remedy:
1. Check operability of the upstream condensate separation device, and rectify cause.
2. Empty, clean and dry the upstream sample gas pipe and sample gas conditioning units.
3. Empty, clean and dry the condensate monitor.
4. Replace filter diaphragm.
5. Press reset switch on the front panel to deactivate the condensate lock.
FAQ
Q: Can SCC-F 23212-0-110310 be installed directly in flammable and explosive areas?
A: The SCC-F 23212-0-110310 standard model is not suitable for potentially explosive atmospheres and must be installed in a non-explosion-proof safe area.
Q: How do I adjust the sample gas flow rate of SCC-F 23212-0-110310?
A: Adjust precisely from 10-100 l/h using the needle valve on the panel in conjunction with the rotor flow meter.
Q: What happens if condensate enters the sample gas path of SCC-F 23212-0-110310?
A: The condensate monitor will trigger an alarm and shut down the diaphragm pump to prevent liquid from entering the expensive analyzer's precision sensors.
Q: Can SCC-F 23212-0-110310 supply two analyzers simultaneously?
A: Yes, this model has two completely independent sample gas paths, supporting multi-channel analysis.
Q: What are the installation location requirements for SCC-F 23212-0-110310?
A: Indoor installation is recommended, and the closer to the analyzer, the better, to minimize response time.