Siemens 7NG3213-0NN00 SITRANS T3K PA Head-Mounted Temperature Transmitter
Manufacturer: Siemens
Condition: Brand New
Lead Time: In Stock
Origin: Germany
Warranty: 12 Months
Email: [email protected]
whatsApp: +86-18150117685
- Overview
- Hot Selling Models
- Recommended Products
Overview
- Do not use the non-Ex article where an explosion-protected transmitter is required.
- Provide a 9–32 V DC PA bus supply for non-Ex operation.
- Match the selected RTD, thermocouple, resistance or millivolt input.
- Use equal lead resistances for three-wire RTD measurements.
- Allocate a unique PA device address during commissioning.
- Select the terminal housing material and protection for the installation’s environmental requirements.
7NG3213-0NN00 SITRANS T3K PA Head-Mounted Temperature Transmitter
A bus-powered, electrically isolated SITRANS T3K PA transmitter for head installation, supplied in the non-Ex version.
At a Glance
| Brand | Siemens |
|---|---|
| Model | 7NG3213-0NN00 |
| Product Type | Temperature measuring transmitter |
| Core Function | Converts sensor measurements to digital process data |
| Application | PROFIBUS PA temperature instrumentation |
Sensor flexibility
Accepts resistance thermometers, thermocouples, resistance and millivolt signals.
PA communication
Transfers measurement and status through PROFIBUS PA.
Head installation
Fits the specified terminal-head mounting arrangement.
Product Overview
7NG3213-0NN00 is the non-explosion-protected SITRANS T3K PA transmitter for installation in a sensor terminal head.
The bus-powered transmitter supports configurable sensor types and standard, averaging or differential measurements, with electrical isolation between the sensor input and PA output.
Key Functions
Configurable temperature acquisition
Supported resistance thermometers and thermocouples can be configured for standard or two-channel averaging and differential measurement.
Resistance and millivolt conversion
Selectable resistance and millivolt input ranges extend the transmitter to compatible non-temperature signals.
Digital fieldbus integration
PROFIBUS PA communication carries cyclic measurement/status information and acyclic parameter access, including class-2 master connections.
Official Specifications
| Non-Ex bus supply voltage | 9–32 V DC, bus-powered |
|---|---|
| Device current consumption | Less than 11 mA |
| Maximum additional fault current | 3 mA |
| Input/output isolation test | 500 V AC, 50 Hz, 1 minute |
| RTD wiring | 2-, 3- or 4-wire for one sensor |
| RTD measurement current | ≤0.55 mA |
| Resistance ranges | 9 selectable ranges from 0–24 Ω to 0–6000 Ω |
| Millivolt ranges | 7 selectable ranges from −1…16 mV to −120…1000 mV |
| Millivolt input resistance | ≥1 MΩ |
| Maximum input overload | 3.5 V |
| Measurement units | Celsius, Kelvin, Fahrenheit or Rankine |
| Class-2 master connections | 4 |
| Factory device address | 126 |
| Long-term drift | Less than 0.1% per year |
| Supply influence on span | Less than 0.005%/V of full-scale range |
| Operating temperature | −40 to +85 °C |
| Storage temperature | −40 to +95 °C |
| Relative humidity | 0–98%, condensing |
| Connection conductor cross-section | Up to 2.5 mm² |
| Housing and weight | PA6 polyamide, GF20; 250 g |
Frequently Asked Questions
Is 7NG3213-0NN00 the explosion-protected T3K PA version?
The 0NN00 article is the version without explosion protection.
Can 7NG3213-0NN00 work with a four-wire resistance thermometer?
A single resistance thermometer can be connected in a two-, three- or four-wire circuit.
How do I avoid line-resistance error with a three-wire sensor on 7NG3213-0NN00?
Use equal lead resistances between the sensor connections and the corresponding transmitter terminals.
Can 7NG3213-0NN00 calculate a temperature difference between two sensors?
The transmitter supports parameterizable differential measurement between two compatible sensor channels.
Does 7NG3213-0NN00 need a separate power pair in addition to PROFIBUS PA?
The transmitter receives its operating power from the PA bus.
Which software can I use to configure 7NG3213-0NN00?
Use SIMATIC PDM for the supported transmitter parameterization functions.
Which address should I expect before commissioning 7NG3213-0NN00?
The transmitter is delivered with device address 126.